ENCORE MEDICAL LP DISCOVERY ELBOW; DISC ULNA 2.5X53MM RT W/ E+ BRNG Back to Search Results
Model Number 540-62-053
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Insufficient Information (4580)
Event Date 06/29/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - due to ulnar component loosening.Put new cement mantle and longer stem in.
 
Manufacturer Narrative
The reason for this revision surgery was reported as loosening.The previous surgery and the surgery detailed in this event occurred 8.3 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA 2.5X53MM RT W/ E+ BRNG
Manufacturer (Section D)
ENCORE MEDICAL LP
9800 metric blvd
austin TX 78758 5445
MDR Report Key12159668
MDR Text Key261310876
Report Number1644408-2021-00703
Device Sequence Number1
Product Code JDC
UDI-Device Identifier00190446254348
UDI-Public(01)00190446254348
Combination Product (y/n)N
PMA/PMN Number
K090473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number540-62-053
Device Catalogue Number540-62-053
Device Lot Number817W1001
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/29/2021
Initial Date FDA Received07/13/2021
Supplement Dates Manufacturer Received07/29/2021
Supplement Dates FDA Received08/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
540-00-004 LOT 75737; 600-10-100 LOT 844E1D0018; 540-00-004 LOT 75737
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-