Model Number AF02
Device Problems Break (1069); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Report confirmed.Evaluation determined that the foot was worn/scraped and the tip was missing.The device was tested and the temperature was measured to be 124.7°f.It was also noted that the device was noisy, the color ring and markings were deteriorated.Previous investigation performed under pr#127716 determined that the likely causes are debris in the collet and improper insertion of the tool.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.In addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.The preventive maintenance/service manual for the legend system specifies service intervals for devices based on the hospital usage level.The maximum specified service interval is 24 months.Device has been in use for approximately 137 months with no record of factory service during this period.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Repair request initiated for device with the report of tip guard breakage.No patient impact reported.Repair escalated to product event due to analysis finding of foot was scraped and the tip was missing.On follow-up, it was confirmed that there was no patient or staff impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX LEGEND FOOTED ATTACHMENT
Type of Device
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
david gustafson
4620 north beach street
ft. worth, TX 76137-3291
7635149628
MDR Report Key12160994
MDR Text Key263460401
Report Number1625507-2021-00244
Device Sequence Number1
Product Code HBB
UDI-Device Identifier00721902481857
UDI-Public00721902481857
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K020069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/24/2015
Device Model NumberAF02
Device Catalogue NumberAF02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Initial Date Manufacturer Received 12/25/2020
Initial Date FDA Received07/13/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-