Model Number 71364022
Device Problems Break (1069); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2021
Event Type  malfunction  
Event Description
It was reported that, during field inspection, it was noticed that the offset ref cup pos/impactor was damaged.No case involved; therefore, there was no patient involvement.
 
Manufacturer Narrative
H6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured and the fractured piece was returned, rendering the device inoperable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OFFSET REF CUP POS/IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12166389
MDR Text Key261471995
Report Number1020279-2021-05953
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010479129
UDI-Public03596010479129
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364022
Device Catalogue Number71364022
Device Lot Number19CM04131
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Initial Date Manufacturer Received 07/06/2021
Initial Date FDA Received07/14/2021
Supplement Dates Manufacturer Received07/30/2021
Supplement Dates FDA Received08/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-