• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STANMORE IMPLANTS WORLDWIDE DISTAL HUMERUS - ELBOW BUSHES; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STANMORE IMPLANTS WORLDWIDE DISTAL HUMERUS - ELBOW BUSHES; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number UNK_STM
Device Problem Naturally Worn (2988)
Patient Problems Ambulation Difficulties (2544); Insufficient Information (4580)
Event Date 06/23/2021
Event Type  Injury  
Manufacturer Narrative
The reported device is similar to a device approved for compassionate use in the united states.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned.
 
Event Description
A patient specific implant form was received for the patient's right distal humerus and elbow prothesis.Reason for surgery is noting: "possible bushing failure at elbow articulation.".
 
Manufacturer Narrative
Reported event: an event regarding rebushing involving a patient specific, distal humeral replacement, bushing was reported.The event was confirmed by x ray review.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: a review of the provided x-rays by a clinical consultant indicated: the implant in situ was for patient specific distal humeral replacement which was inserted on (b)(6) 2014.The surgeon reported a possible bushing failure at elbow articulation.The lateral view x-ray shows a gap at the elbow articulation between the humeral and ulnar components, which indicates a loosed elbow joint.A possible reason for this is wear of the bush.Therefore, the radiographic review can confirm the clinical report and reason for revision.Device history review: review of the product history records indicate the device was manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for the lot referenced.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, pathology reports, progress notes, and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event Description
A patient specific implant form was received for the patient's right distal humerus and elbow prothesis.Reason for surgery is noting: "possible bushing failure at elbow articulation.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAL HUMERUS - ELBOW BUSHES
Type of Device
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK  WD6 3SJ
Manufacturer (Section G)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK   WD6 3SJ
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key12171723
MDR Text Key261663792
Report Number3004105610-2021-00111
Device Sequence Number1
Product Code KWS
Combination Product (y/n)N
Reporter Country CodeGR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Catalogue NumberUNK_STM
Device Lot NumberPIN 19126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/23/2021
Initial Date FDA Received07/15/2021
Supplement Dates Manufacturer Received11/26/2021
Supplement Dates FDA Received12/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age36 YR
Patient SexFemale
-
-