ABBOTT LABORATORIES ACCELERATOR APS CENTRIFUGE MODULE; CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM Back to Search Results
Catalog Number 07L02-53
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/12/2021
Event Type  malfunction  
Manufacturer Narrative
A product deficiency was identified due to an increase in complaints for failures of the r100 component of the main electronics board (parts (b)(4)).The supplier, inpeco, determined that the issue is from the r100 component of the hettich centrifuge board which was too small to withstand a current spike.Hettich released an enhanced main board revision 06 with a different r100 resistor.It is considered acceptable to continue using the automation system because the issue does not impact operator nor patient safety.The accelerator aps operations manual and the centrifuge operations manual were reviewed and show that troubleshooting and safety information is provided.There have been no reported injuries for this issue.If the electrical components of the main electronics board fail, it will shut down the centrifuge module.If the centrifuge shuts down, the operator will receive an error message.The frequency of this hazard was assessed and it was determined that though there was a trend in complaints, the overall frequency of low had not changed from the predicted frequency as outlined in the risk management file.
 
Event Description
The customer is unable to get the centrifuge module to power on.There was no smoke or fire observed.There was no impact to patient management, user safety, or facility damage reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCELERATOR APS CENTRIFUGE MODULE
Type of Device
CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM
Manufacturer (Section D)
ABBOTT LABORATORIES
1915 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT LABORATORIES
1915 hurd drive
irving TX 75038
Manufacturer Contact
christian lee
100 abbott park road
dept. 09b9, lc cp01-3
abbott park, IL 60064-3537
2246682940
MDR Report Key12173687
MDR Text Key261709164
Report Number3016438761-2021-00256
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07L02-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/29/2021
Initial Date FDA Received07/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-