Model Number OER-PRO
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2021
Event Type  malfunction  
Manufacturer Narrative
The olympus field service engineer (fse) replaced the hoses.The fse will complete the repair when a check valve is received.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Event Description
It was reported by an olympus field service engineer (fse) that the hoses in the endoscope reprocessor were degrading.This was causing black specks in the disinfectant.It was reported this was found during preventative maintenance.There was no patient harm or impact to patient care due to this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the investigation, the inner wall of the hose inside the equipment was peeling off in fragments, which came out as black foreign material during reprocessing the reporters occupation is director of surgical services.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12174358
MDR Text Key261746069
Report Number8010047-2021-08908
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/18/2021
Initial Date FDA Received07/15/2021
Supplement Dates Manufacturer Received09/24/2021
Supplement Dates FDA Received09/29/2021
Is This a Reprocessed and Reused Single-Use Device? No