Of the 10 impactors shipped 10/9/2020 by orthogroup, this is the third straight threaded impactor that had broken distal tips upon impact.The event on (b)(6) 2021 was the only one that caused a significant delay in surgery.Although these instruments are designed and manufactured primarily by orthogroup, nextstep is evaluating these impactors to try and replicate the issue.There is significant suspicion that improper assembly of the cup onto the impactor (not fully seated) is contributing to the breakage, as there is evidence of wear on the bow-tie edges that are intended to seat in provided detents.
|