• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST SINGLE DETECTOR; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST SINGLE DETECTOR; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Back to Search Results
Model Number 712020
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Int.Ref: (b)(4).The digitaldiagnost single detector is a direct digital radiography system with flat detector technology based on modular components to allow for customization for all radiographic applications and workload requirements.For anatomies that are larger than the detector size, it is possible to make a series of exposures covering the whole anatomy (for example full spine or full legs).These individual images can be "stitched" together via the software program.For proper patient positioning and support, the patient can be placed on the so called stitching patient support or patient support for stitching (9890 010 87431), a platform with two handles that the patient can hold on to during the examination.For the ease of use during transportation, the patient support for stitching has wheels and a folding footboard.The footboard is connected to the frame via two hinges and a brake cylinder.It has to be folded up and fixed by a hook for transportation, e.G.From one room to another.The brake cylinder has the task to ensure that the footboard lowers itself slowly (within 5-7 seconds) when the hook is released, to prevent the footplate from falling on the foot of a person.If the operator steps on the footboard while it is still moving down, the mounting of the brake cylinder can break and the footboard falls down immediately.In a worst case, it may hit the foot of a person and break a bone.The philips healthcare field service engineer (fse) has investigated at site and found the frame damaged (cast aluminum was broken).The cause was not identified but most probable a person stepped onto the footplate to force down movement before it was fully released on the floor.The fse has replaced the patient support for stitching.System works as specified again.A capa investigation was conducted with the result of acceptable risk per benefit risk determination.This issue is further monitored and trended.
 
Event Description
The customer complained that the mounting of the brake cylinder at the patient support for stitching failed.No patient or user harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST SINGLE DETECTOR
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM  22335
MDR Report Key12175680
MDR Text Key261780192
Report Number3003768251-2021-10010
Device Sequence Number1
Product Code MQB
Combination Product (y/n)N
PMA/PMN Number
K090625
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712020
Device Catalogue Number712020
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date04/14/2021
Initial Date Manufacturer Received 04/14/2021
Initial Date FDA Received07/15/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-