• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INJECTION GOLD PROBE; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INJECTION GOLD PROBE; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00560150
Device Problems Crack (1135); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental emdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an injection gold probe was used in the stomach during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2021.During the procedure, the physician advanced the gold probe into the biopsy cap and endoscope.The device would not work once the tip exited the scope, so the staff withdrew the device from the scope.The distal gold tip detached from the end of the catheter and remained in the medivator's biopsy cap.In addition, the distal gold tip was cracked.The procedure was completed with a different device which was not specified.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INJECTION GOLD PROBE
Type of Device
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12177860
MDR Text Key261896553
Report Number3005099803-2021-03513
Device Sequence Number1
Product Code KNS
UDI-Device Identifier08714729056492
UDI-Public08714729056492
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133933
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/23/2022
Device Model NumberM00560150
Device Catalogue Number6015
Device Lot Number0025630861
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received07/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-