• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE Back to Search Results
Model Number EG29-I10C
Device Problems Leak/Splash (1354); Defective Component (2292); Material Deformation (2976); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the defect was returned.We confirmed that the pve connector leak and the suction channel buckled, the ift bump, and u/d and r/l pulley wire worn out based on our investigation of the returned defect.This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c-us is available in the usa with a 510k number k190805.This report is being filed as part of the pentax backlog management plan.
 
Event Description
This event occurred during inspection.There was no report of patient harm.Led board defective, pve connector leaky - with test pcb the led is ok.Sold date (b)(6) 2019, last repair 2021-03-12.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
IMAGINA GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
0804315880
MDR Report Key12178904
MDR Text Key261944337
Report Number9610877-2021-10090
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333248483
UDI-Public04961333248483
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/27/2021
Initial Date FDA Received07/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-