Model Number CLV-190
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/17/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user report "front display was acting up" was confirmed.The lamp hours were not displaying on the lamp indicator due to a faulty main printed circuit board.Consequently, the light output measured outside of the specifications.The scope socket condition was ok, the main power switch was up to date, the fan was functioning properly, and the air pressure tested within range.However, there was corrosion noted inside the air tubing.Furthermore, the turret filter was faulty and the lamp will require replacing.If additional information becomes available following device evaluation, a supplemental report will be filed.
 
Event Description
The customer called in to report that the light source on the evis exera iii xenon light source began smoking during a procedure and the image became distorted.According to the initial reporter, they inspected the lamp to see why it was smoking, but it appeared to be fine.The user decided to switch to another tower to successfully complete the procedure.There was no patient harm or consequence reported as a result of this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12181488
MDR Text Key261986130
Report Number8010047-2021-08971
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/17/2021
Initial Date FDA Received07/16/2021
Supplement Dates Manufacturer Received01/27/2022
Supplement Dates FDA Received02/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-