Model Number CATRXKIT
Device Problem Suction Failure (4039)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure in the left anterior descending artery (lad) using an indigo system catrx aspiration catheter (catrx), a penumbra engine (engine), a penumbra engine canister (canister), a non-penumbra rhv and a guidewire (.014 wire).During the procedure, the physician completed the first pass using the catrx.The physician then advanced the catrx to the area of clot for the second pass and turned the aspiration on; however, there was no flow and the catrx was unable to aspirate the thrombus.He also tried to troubleshoot the canister and the engine but found no issue with them.All four lights of the engine were illuminated.Therefore, the physician decided to remove the catrx.It was reported that there was no damage seen on the catrx after it was removed.The procedure was completed using another catrx.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM CATRX ASPIRATION CATHETER
Type of Device
QEX, QEW
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12186539
MDR Text Key262198276
Report Number3005168196-2021-01581
Device Sequence Number1
Product Code QEX
UDI-Device Identifier00814548017556
UDI-Public00814548017556
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K163618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,07/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/12/2024
Device Model NumberCATRXKIT
Device Catalogue NumberCATRXKIT
Device Lot NumberF103864
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 06/22/2021
Initial Date FDA Received07/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
-
-