• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The manufacturing record was reviewed and found no irregularities.Omsc checked the subject device and found that the reported phenomenon could not be duplicated.Omsc checked the error log and found that e03 (pressure sensor abnormality) had occurred.Therefore, omsc presumed that the reported phenomenon occurred due to the abnormality in the pressure sensor.
 
Event Description
Olympus medical systems corp.(omsc) was informed that at the end of the laparoscopic inguinal hernia repair, the abdominal pressure dropped to 2 mmhg and there was almost no flow rate.After that, an alarm sounded and the supply stopped.The procedure was completed without replacing the equipment.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12187249
MDR Text Key262516155
Report Number8010047-2021-09038
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/30/2021
Initial Date FDA Received07/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-