• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN LEGION TOTAL KNEE PRIM FEM COMP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN LEGION TOTAL KNEE PRIM FEM COMP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number UNKN01200212
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Inadequate Osseointegration (2646); Joint Laxity (4526)
Event Date 05/22/2018
Event Type  Injury  
Event Description
It was reported that, after a tka surgery was performed on (b)(6) 2016, the patient experienced asepting femoral component loosening with tka instability.A revision surgery had to be conducted on (b)(6) 2018.The revised device was an unknown legion total knee prim fem comp.The site indicated that no supporting source documents can be submitted, other than the ones already provided.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a right tka revision was required approximately 2½ years post implantation due to aseptic loosening of the femoral component with tka instability.It was communicated that ¿no supporting source documents can be submitted, other than the ones already provided¿.Event was resolved on (b)(6) 2018.The provided case report forms did not provide further insight into the root cause of the reported events.Patient impact beyond the reported loosening, instability, and subsequent revision could not be determined.No further medical assessment could be rendered at this time.Should clinically relevant documentation become available in the future, this case may be re-opened/re-assessed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as abnormal motion over time, fit/sizing issue, lifetime of device, patient condition and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN LEGION TOTAL KNEE PRIM FEM COMP
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12201276
MDR Text Key262609640
Report Number1020279-2021-06015
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,s
Type of Report Initial,Followup
Report Date 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKN01200212
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/09/2021
Initial Date FDA Received07/21/2021
Supplement Dates Manufacturer Received08/25/2021
Supplement Dates FDA Received08/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight62
-
-