Catalog Number ASKU |
Device Problem
Filling Problem (1233)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the unspecified access set under-infused.The device was filled with argatroban while being used with spectrum iq infusion pump.The customer stated that the only half of the dose was received by the patient over two days due to clip placement.It was further reported that the nurse did not fully open the blue slide clamp on the iv tubing on the spectrum pump; nor did they check to make sure the blue clip was entirely loose from the iv tubing.This issue was identified during patient use.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|