Device Problems Migration or Expulsion of Device (1395); Material Separation (1562)
Patient Problems Pulmonary Edema (2020); Cardiogenic Shock (2262); Foreign Body Embolism (4439); Mitral Valve Insufficiency/ Regurgitation (4451)
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
The article, "emergency valve re-replacement for embolization of prosthetic mitral valve disc during catheterization procedure", was reviewed.This research article presents a case study on a (b)(6) women with rheumatic heart disease, who underwent tissue mitral valve replacement remotely, which was replaced with a mechanical prosthesis in 1996 due to structural valve deterioration.In 2002, the patient had an emergency valve re-replacement with a st.Jude bi-leaflet mechanical valve for acute valve thrombosis during pregnancy.The patient recently experienced symptomatic paroxysmal atrial fibrillation, for which the patient underwent catheter ablation.During the procedure the patient experienced acute severe mitral regurgitation.This was visualized on fluoroscopy as coinciding with the dislodgment of one of the mechanical valve leaflets, which embolized to the descending thoracic aorta.The patient immediately exhibited pulmonary edema and cardiogenic shock and was supported with a left ventricular assist device.The patient was transferred to the operating room to undergo emergency surgery.The patient underwent re-operative sternotomy and became hemodynamically unstable with acute right ventricular failure after the repeat sternotomy.Cardiopulmonary bypass was initiated.Under cold cardioplegic arrest, the mechanical mitral valve was exposed through the interatrial groove.One of the leaflets of the valve was confirmed to be missing, but the remainder of the prosthesis was intact.The damaged prosthesis was replaced with a 27mm non-abbott mechanical valve.The patient was extubated the following day and transferred to the ward on the third postoperative day.Computed tomographic angiography on day 3 located the missing disc in the abdominal aorta.On postoperative day 6 the patient underwent a laparotomy with the distal abdominal aorta exposed through a retroperitoneal approach and the leaflet was extracted.The patient was discharged home uneventfully one week after the abdominal surgery.The patient remained well, asymptomatic, and in sinus rhythm a year after the procedure.The article concluded that with a growing cohort of patients with mechanical mitral prostheses and expanding utilization of percutaneous interventions for atrial arrhythmias and also for transcatheter valve replacement and treatment of paravalvular leaks, surgeons and interventional cardiologists should be familiar with the potential for this complication and its management.The primary author of the article is amit pawale, md, department of cardiothoracic surgery, and division of vascular surgery, department of surgery, mount sinai medical center, new york, new york.The correspondence author of the article is anelechi c.Anyanwu, md, department of cardiothoracic surgery, mount sinai medical center, 1190 fifth ave, gp 2w, box 1028, new york, ny 10029 with the corresponding email: anelechi.Anyanwu@mountsinai.Org.
 
Manufacturer Narrative
As reported in a research article, a patient had an emergency valve replacement due to thrombosis and one of the valve leaflets embolizing during a catheter ablation procedure.A more comprehensive assessment could not be performed as the event was non-contemporaneously reported through a literature review and no device was received for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key12217425
MDR Text Key263134271
Report Number2648612-2021-00079
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,literatur
Type of Report Initial,Followup
Report Date 07/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/24/2021
Initial Date FDA Received07/23/2021
Supplement Dates Manufacturer Received07/27/2021
Supplement Dates FDA Received07/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age47 YR
-
-