The reason for this complaint was reported as damage incurred from prolonged use and through misuse of instrument.The possible time in service for baseplate is 9 months from manufactured date.The healthcare professional indicated that this event occurred during surgery, near the patient.No risk, adverse event, or negative outcome were reported by the surgeon.The surgery was completed as intended, with a five-minute delay.The instrument was inspected prior to use and was deemed acceptable for use based on its appearance.The agent was present during surgery but was unable to source a suitable replacement device.The device was returned to manufacturer and evaluated by registered medical assistant (rma) djo surgical.A review of the instrument's device history record (dhr) revealed the instrument, when released for use, met design and manufacturing requirements.There were no ncmr's associated with the production of the instruments that are related to the reported issue.Complaint database review shows no prior complaints filed against the instrument's production lot that report a similar failure.No prior complaints report past instances of these instruments being affected by this failure.The root cause of this complaint is likely due to damage incurred from prolonged use and through misuse or rough handling which surgical instruments are subjected to.There are no indications that this instrument has a systemic design or material deficiency.Therefore, no containment of inventory is required.Event is associated with instrument usage, not a design or manufacturing issue.Rma examination: the reported instrument was returned to djo in one piece.Upon further examination a major crack was found and the djo empowr kneetm, punch handle is actually in two pieces.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|