MEDTRONIC EUROPE SARL CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTMA1D1 |
Device Problem
Misconnection (1399)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/21/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: 693565 lead, implanted: (b)(6) 2021.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the right ventricular (rv) lead alerted for high/undefined pacing impedance.No capture was also observed.There was a loose connection between the lead and the cardiac resynchronization therapy defibrillator (crt-d).The lead was removed from the device header, tested, and reconnected to the device.The lead and device remain in use.No patient complications have been reported as a result of this event.
|
|