Model Number CLV-S190 |
Device Problem
Energy Output Problem (1431)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/02/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was not returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the preparation for use, the customer found xenon lamp didn't get power on and it was shifted to spare lamp.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This is a supplemental report to provide additional information.The subject device was returned to local service department of olympus.Local service department checked the subject device and found that the reported phenomenon was duplicated due to the failure of power supply unit.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Omsc presumed that reported phenomenon occurred due to the converter failure.The cause of the failure might be deterioration due to repeated use because 4 years had passed since production, or accidental failure.
|
|