• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AK 96; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AK 96; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114364
Device Problem Infusion or Flow Problem (2964)
Patient Problems High Blood Pressure/ Hypertension (1908); Swelling/ Edema (4577)
Event Date 06/03/2021
Event Type  Injury  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that, during treatment with hemodialysis machine ak 96, air and water leakage occurred at the elbow on negative pressure pump.After 3 hours and 40 minutes of dialysis treatment, the patient felt swelling and discomfort of the upper limbs and face.The treatment was stopped.The accumulated ultrafiltration volume displayed by the ak 96 showed 2.19l.Post treatment weight recording was (b)(6) kg indicating that the patient had gained 3.7l compared to the pre treatment weight and their blood pressure had increased to 183/90 mmhg.No alarm was generated during the treatment.A new 4 hour treatment was initiated, using a different ak 96 dialysis machine and the patient reportedly had no discomfort.No additional information is available.
 
Manufacturer Narrative
Additional information added to h6 and h10.H10: the device was not received for evaluation; therefore, a device analysis could not be completed.A device history review revealed no issues that could have caused or contributed to the reported issue.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12224734
MDR Text Key263392970
Report Number9616026-2021-00036
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number114364
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/29/2021
Initial Date FDA Received07/26/2021
Supplement Dates Manufacturer Received08/09/2021
Supplement Dates FDA Received08/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age49 YR
Patient Weight56
-
-