Model Number BB515
Device Problem Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2021
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with bb515 - carbon steel scalpel blades #15.According to the complaint description, the blade was losing the cutting wire very fast.There was no described patient harm.Additional information was not provided.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Updated batch number and expiration date: h4: updated device manufacturer date: h6: updated codes ; investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBON STEEL SCALPEL BLADES #15
Type of Device
SCALPEL BLADES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12225635
MDR Text Key263433977
Report Number9610612-2021-00518
Device Sequence Number1
Product Code GES
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Model NumberBB515
Device Catalogue NumberBB515
Device Lot Number4509219957
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/06/2021
Initial Date FDA Received07/26/2021
Supplement Dates Manufacturer Received08/17/2021
Supplement Dates FDA Received09/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-