MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problems Loss of Power (1475); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that during surgery system lost power and completely shut off.The procedure was completed and patient harm was not reported.Additional information has been requested.
 
Manufacturer Narrative
The company service representative, examined the system.But did not replicate the reported event.There was no indication of power loss.And it was in great condition.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The system was found to meet specifications.Therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key12233136
MDR Text Key263882252
Report Number2028159-2021-00813
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657511501
UDI-Public00380657511501
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/01/2021
Initial Date FDA Received07/27/2021
Supplement Dates Manufacturer Received09/10/2021
Supplement Dates FDA Received10/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-