• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPACEOAR VUE SYSTEM; ABSORBABLE PERIRECTAL SPACER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPACEOAR VUE SYSTEM; ABSORBABLE PERIRECTAL SPACER Back to Search Results
Model Number SV-2101
Device Problems Obstruction of Flow (2423); Material Integrity Problem (2978); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that spaceoar vue was attempted to be implanted during a spaceoar vue placement procedure performed (b)(6) 2021.The procedure was performed under general anesthesia.During the procedure, there was significant difficulty finding proper hydrodissection due to axial imaging fault and needle visualization.Once hydrodissection was appropriate, an attempt to inject gel was made but appeared to have been clogged at y-connector.Procedure was aborted and no gel was placed.Needle also appeared to be bent when removed from patient.The patient was already sedated and the procedure was cancelled.There were no patient complication reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEOAR VUE SYSTEM
Type of Device
ABSORBABLE PERIRECTAL SPACER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12233160
MDR Text Key263854444
Report Number3005099803-2021-03826
Device Sequence Number1
Product Code OVB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182971
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/20/2022
Device Model NumberSV-2101
Device Catalogue NumberSV-2101
Device Lot Number0026743771
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/30/2021
Initial Date FDA Received07/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-