ENCORE MEDICAL L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 7R 10MM, VE Back to Search Results
Model Number 342-10-707
Device Problem Material Too Rigid or Stiff (1544)
Patient Problem Loss of Range of Motion (2032)
Event Date 07/15/2021
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as stiffness.The previous surgery and the surgery detailed in this event occurred 4 years and 8 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to stiffness.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - the patient developed stiffness over the course of 5 years.The patient's range of motion was 0-40 prior to revision surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR 3D KNEE
Type of Device
EMPOWR 3D KNEETM INS, 7R 10MM, VE
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key12234670
MDR Text Key263842114
Report Number1644408-2021-00773
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00888912167345
UDI-Public(01)00888912167345
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/26/2021
Device Model Number342-10-707
Device Catalogue Number342-10-707
Device Lot Number077T1028
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/15/2021
Initial Date FDA Received07/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient Weight99
-
-