Model Number CLV-190
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at local service department of olympus on july 8, 2021, it was found that moisture had penetrated the air hose.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.According to the evaluation result from the local service department, it was confirmed that there was a sign that the liquid had infiltrated the tube, so it was considered that the user inserted the scope without thoroughly wiping the moisture after washing the endoscope and the output connector.It was also considered that there might be a problem with the endoscope causing liquid backflow, or a defective operation of the pump of the subject device, a part failure such as a reverse valve, or an accidental failure.In addition, because rust was found outside of the device, it was also considered that it infiltrated into the tube due to the effect of invading the inside by mistake.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12237220
MDR Text Key280956492
Report Number8010047-2021-09458
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2021
Initial Date FDA Received07/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-