Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at local service department of olympus on july 8, 2021, it was found that moisture had penetrated the air hose.There was no report of patient injury associated with the event.
|
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.According to the evaluation result from the local service department, it was confirmed that there was a sign that the liquid had infiltrated the tube, so it was considered that the user inserted the scope without thoroughly wiping the moisture after washing the endoscope and the output connector.It was also considered that there might be a problem with the endoscope causing liquid backflow, or a defective operation of the pump of the subject device, a part failure such as a reverse valve, or an accidental failure.In addition, because rust was found outside of the device, it was also considered that it infiltrated into the tube due to the effect of invading the inside by mistake.
|