Model Number CDS0701-NTW |
Device Problem
Difficult to Flush (1251)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report a difficult to flush.It was reported that the clip delivery system (cds) was prepared without issue; however, when the cds was being moved to the table, it was noted that one of the drip lines was not working which indicates the cds could not be properly flushed.The cds was not used in the patient and the procedure was successfully completed with a new cds.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The returned device analysis did not confirm the reported difficult to flush.A review of the lot history identified no manufacturing nonconformities issued to the reported lot that would have influenced the reported issue.Additionally, a review of the complaint history did not identify any similar incidents.All available information was investigated, a cause for the reported difficult to flush cannot be determined in this incident.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|
Event Description
|
N/a.
|
|