• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-NTW
Device Problem Difficult to Flush (1251)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2021
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report a difficult to flush.It was reported that the clip delivery system (cds) was prepared without issue; however, when the cds was being moved to the table, it was noted that one of the drip lines was not working which indicates the cds could not be properly flushed.The cds was not used in the patient and the procedure was successfully completed with a new cds.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The returned device analysis did not confirm the reported difficult to flush.A review of the lot history identified no manufacturing nonconformities issued to the reported lot that would have influenced the reported issue.Additionally, a review of the complaint history did not identify any similar incidents.All available information was investigated, a cause for the reported difficult to flush cannot be determined in this incident.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12239786
MDR Text Key263918288
Report Number2024168-2021-06593
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648230967
UDI-Public08717648230967
Combination Product (y/n)N
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/10/2022
Device Model NumberCDS0701-NTW
Device Catalogue NumberCDS0701-NTW
Device Lot Number10111U129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2021
Initial Date Manufacturer Received 07/13/2021
Initial Date FDA Received07/28/2021
Supplement Dates Manufacturer Received09/13/2021
Supplement Dates FDA Received09/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-