BECTON DICKINSON, S.A. SYRINGE 1ML LS SP120; PISTON SYRINGE Back to Search Results
Catalog Number 303172
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the syringe 1ml ls sp120 was difficult to connect to the tfk laboratory needle.The following information was provided by the initial reporter, translated from (b)(6) to english: "this is a report about syringe needle connectivity issue.The customer uses this product for covid vaccination." "the connection of this syringe (cat#303172) to tsk laboratory¿s needle became loose in 20 of 320 products.".
 
Manufacturer Narrative
H.6.Investigation: twenty two samples and photos received for investigation, upon visual inspection of the samples received, no molding or other defect can be noticed in the syringe, no burrs or deformations can be seen in the tip.Additionally, ten retained samples of lot 200612 were used to conduct a leakage test.The product was visually inspected, no defects or damage was noted, and no leak was identified.A device history review was performed for lot 2006012, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Based on the available information we are not able to determine a root cause at this time.
 
Event Description
It was reported that the syringe 1ml ls sp120 was difficult to connect to the tfk laboratory needle.The following information was provided by the initial reporter, translated from japanese to english: "this is a report about syringe needle connectivity issue.The customer uses this product for covid vaccination." "the connection of this syringe (cat#303172) to tsk laboratory¿s needle became loose in 20 of 320 products.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1ML LS SP120
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
MDR Report Key12242750
MDR Text Key264726590
Report Number3003152976-2021-00432
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number303172
Device Lot Number2006012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2021
Initial Date Manufacturer Received 06/29/2021
Initial Date FDA Received07/28/2021
Supplement Dates Manufacturer Received09/13/2021
Supplement Dates FDA Received09/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-