• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. SHELL 49 MM O.D.; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. SHELL 49 MM O.D.; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial hip arthroplasty, the locking ring was dislodged from the cup.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified lock ring is not correctly oriented in lock ring groove.Lock ring tab was stuck in groove and not centered in tab slot window.Damage was noted to the lock ring upon arrival.Lock ring was removed for evaluation and shows deformation and damage.No other damage to the device was noted.Complaint sample was evaluated and the reported event was not confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHELL 49 MM O.D.
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12249195
MDR Text Key264266426
Report Number0002648920-2021-00193
Device Sequence Number1
Product Code KWY
UDI-Device Identifier00889024115415
UDI-Public(01)00889024115415(17)310323(10)62344694
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K833991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model NumberN/A
Device Catalogue Number00500104900
Device Lot Number62344694N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/06/2021
Initial Date FDA Received07/29/2021
Supplement Dates Manufacturer Received11/16/2021
Supplement Dates FDA Received11/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-