• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII LONG I/M ROD 16 INCH; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII LONG I/M ROD 16 INCH; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440004
Device Problems Break (1069); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2021
Event Type  malfunction  
Event Description
It was reported that, during a field inspection, these devices were found with issues: gii long i/m rod 16 inch was found broken and bent (case (b)(4)), a large outer case - 4.8 (case (b)(4)) and a lid for outer cases (case (b)(4)) were found cracked, and a g2 con arti tr sz3-4 11mm (case (b)(4)) was found with burrs, nicks, scratches and/or gouges.No case involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII LONG I/M ROD 16 INCH
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12250730
MDR Text Key264325121
Report Number1020279-2021-06110
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010213723
UDI-Public03596010213723
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/16/2006
Device Model Number71440004
Device Catalogue Number71440004
Device Lot Number60802736
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/15/2021
Initial Date FDA Received07/29/2021
Supplement Dates Manufacturer Received09/14/2021
Supplement Dates FDA Received09/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-