The device was evaluated at olympus (b)(4), and olympus (b)(4) confirmed the following: according to the repair history, the device was repaired on september 17, 2020.There was a leakage from the bending section.The insertion section was yellowish due to aging.Olympus medical systems corp.(omsc) could not confirm any defects that could cause the reported event from the device inspection result.The exact cause of the reported event could not be conclusively determined, because the device has not been returned to omsc.However, based upon the past similar cases, the reported event may have been caused by the application of some physical or chemical stress to the insertion section.Examples include physical stress such as rubbing the insertion section, chemical stress due to deviation from the reprocessing manual, poor storage environment such as direct sunlight, high temperature, high humidity, x-rays and ultraviolet rays, aging deterioration, etc.The instruction manual provides warnings about applying external stress to the insertion section, reprocessing method not recommended by the reprocessing manual, and storage of the endoscope in a harsh environment.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
|