• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Component Incompatible (1108)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Manufacturer Narrative
Device not accessible for testing:.A getinge field service engineer (fse) that encountered the issue, reported that the customer's biomed is trained and will perform the repair.Additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted when this information is provided to us.The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is: (b)(6).
 
Event Description
It was reported that during software update, performed by a getinge field service engineer (fse),the cardiosave intra-aortic balloon pump (iabp) would not pass drive manifold test.There was no patient involvement, and there was no adverse event reported.
 
Manufacturer Narrative
Additional contact information: (b)(6).A getinge field service engineer (fse) that encountered the issue, reported that the customer's biomed is trained and will perform the repair.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12258200
MDR Text Key264461830
Report Number2249723-2021-01673
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/08/2021
Initial Date FDA Received08/02/2021
Supplement Dates Manufacturer Received02/14/2024
Supplement Dates FDA Received02/15/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/05/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-