Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 07/13/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: item number: 01.06010.204, item name: avenir std stem cemented 4 lot # 3064653.The device will not be returned for analysis remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental mdr will be submitted.
|
|
Event Description
|
It was reported that a new avenir complete 12/14 taper 32mm +3.5 head trial disassociated from trunnion by soft tissues and became irretrievable.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records, lot identification was not provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|