Model Number UM-S20-17S
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/08/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed the following information.As a result of visual inspection, the ultrasonic medium was leaked from the insertion tube.There were scratches and indentations on the tip sheath.The ultrasound image was properly displayed only when the bubbles are removed.Serial number: (b)(4) (manufacturing date: mar 3, 2021).Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the investigation results, omsc surmised that the cause of the reported event is as follows.The ultrasonic medium leaked since the tip sheath was scratched and punctured.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user, it was found the leakage of the ultrasonic medium from the insertion tube.The ultrasonic image of the subject device was normally displayed, and there was no apparent abnormality.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12264724
MDR Text Key265620594
Report Number8010047-2021-09693
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/09/2021
Initial Date FDA Received08/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-