• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problems Device Contamination with Chemical or Other Material (2944); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Local service department checked the subject device and found that the phenomenon occurred due to a shock caused by dropping and knocking the endoscope to a hard surface or object.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at local service department of olympus on (b)(6) 2021, gap of adhesive on the distal end and stain entered inside the lens through the gap were found.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Omsc presumed that the adhesive peeled off because an excessive force was applied when handling the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12267232
MDR Text Key264765637
Report Number8010047-2021-09717
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356346
UDI-Public04953170356346
Combination Product (y/n)N
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/16/2021
Initial Date FDA Received08/03/2021
Supplement Dates Manufacturer Received08/25/2021
Supplement Dates FDA Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-