Model Number FHC-A202
Device Problem False Positive Result (1227)
Patient Problem Distress (2329)
Event Type  malfunction  
Manufacturer Narrative
Results pending investigation.
 
Event Description
Unspecified date: patient presented to facility for a planned surgery (type of surgery unspecified).Customer states it was not an emergent situation.A patient urine specimen was tested on the fisher sure-vue hcg stat serum/urine kit and a positive result was obtained.A patient serum specimen was tested on the fisher sure-vue hcg stat serum/urine kit and a negative result was obtained.A confirmatory beta quant hcg test was performed on the beckman coulter dxi analyzer and a negative result of <0.6 miu/ml was obtained.The patients surgery was initially delayed, however, it was performed 3 hours later.Customer states the patient experienced emotional distress, however, no adverse outcomes occurred.No remedial actions were needed.
 
Manufacturer Narrative
D9: device available for evaluation changed to "no" h3: although requested, devices were not returned.Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical urine samples and sub-cutoff urine standards (4 miu/ml).The results were read at 3 and 4 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.Although the tracking number shows that the returned devices were delivered in san diego, a search of the facility did not locate the product.Therefore, testing of returned devices could not be included in the investigation.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert, this test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER-SURE-VUE HCG STAT SRM/URINE (50T)
Type of Device
PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
MDR Report Key12268242
MDR Text Key264778856
Report Number2027969-2021-00074
Device Sequence Number1
Product Code JHI
UDI-Device Identifier20885380020268
UDI-Public(01)20885380020268(17)221231(10)HCG0122141
Combination Product (y/n)N
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberFHC-A202
Device Catalogue NumberB1077-23
Device Lot NumberHCG0122141
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/19/2021
Initial Date FDA Received08/03/2021
Supplement Dates Manufacturer Received08/21/2021
Supplement Dates FDA Received08/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ORTHOTRICYCLEN; ORTHOTRICYCLEN
Patient Age42 YR
Patient Weight96
-
-