• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SP*2 FEMORAL IMPACTOR; KNEE INSTRUMENT : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SP*2 FEMORAL IMPACTOR; KNEE INSTRUMENT : IMPACTORS Back to Search Results
Model Number 96-6180
Device Problem Unintended Movement (3026)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint #
=
> (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that upon impacting the femur with the secondary impactor, the plastic impactor portion came loose and no longer could be held on by the screws of the handle.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Question: a.Please verify if the instrument used during surgery? if yes, was there surgical delay? what was the duration of the delay? answers for your question: yes, 8 minutes to get another tray.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received confirmed that the instrument was used during surgery.There was a surgical delay of 8 minutes.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device did not confirm the reported breakage, but did find deformation.The noted damage is consistent with device wear from normal use and servicing and the investigation did not establish a need for corrective action.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12274289
MDR Text Key264964033
Report Number1818910-2021-16976
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295247203
UDI-Public10603295247203
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96-6180
Device Catalogue Number966180
Device Lot NumberSY0811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Initial Date Manufacturer Received 07/20/2021
Initial Date FDA Received08/04/2021
Supplement Dates Manufacturer Received08/02/2021
08/04/2021
08/13/2021
08/10/2021
09/10/2021
Supplement Dates FDA Received08/05/2021
08/16/2021
08/18/2021
08/24/2021
09/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-