TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's sales associate, the perfusionist reported that the inaccurate pco2 and po2 values were displayed after in-vivo calibration.A second calibration did not correct the trending accuracy.The actual po2 was 47 and the bpm was trending it at 37.
|
|
Event Description
|
It was reported that during use of the device for cardiopulmonary bypass (cpb), the blood parameter monitor (bpm) displayed different partial pressure of oxygen (po2) and partial pressure of carbon dioxide (pco2) values than the blood gas analyzer.As a mitigation, more blood gases were done.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Event Description
|
Per clinical review: the team at the user facility had an incident with the blood parameter monitor (bpm) during a cardiopulmonary bypass (cpb) procedure on (b)(6) 2021.The team set up and calibrated the bpm without issue.After the first in-vivo calibration on cpb, the perfusionist noticed that the partial pressure of carbon dioxide (pco2) and partial pressure of oxygen (po2) values were very off from the blood gases, example of pco2 of 37 millimeters of mercury (mmhg) on the bpm and the blood gas analyzer was reading 47 mmhg.The team did more blood gas values during the procedure, and exchanged the bpm after the procedure.There was no delay in the continuation of the surgical procedure.There was no harm or blood loss.
|
|
Manufacturer Narrative
|
Updated block: h6.The reported complaint was not confirmed.The service repair technician (srt) could not duplicate the issue.The blood parameter monitor (bpm) powered on and passed the self-test.The arterial bpm passed intensity testing, the hematocrit saturation (hsat) passed service mode testing, and the erasable electronically programmable read only memory (eeprom) contained 0 critical errors.The monitor passed all testing.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) put the blood parameter monitor (bpm) into operate mode with a shunt sensor attached to the bpm and a cuvette attached to the hematocrit saturation (hsat).The system was left alone for approximately five minutes, then the store/recall function was used to adjust on-screen values with no issues observed.A sample calibration was successful.The operate mode observation was repeated and no issues observed adjusting on-screen values.The monitor was left in operate mode for two hours and the values did not change unexpectedly during that time.
|
|
Search Alerts/Recalls
|
|
|