Additional information was added to h3, h4, and h6.H4: the device was manufactured from february 24, 2021 - february 25, 2021.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.When the luer cap was removed, evidence of continuous flow of fluid was observed coming out of the line.A functional flow rate test was performed and found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|