Catalog Number UNKN01100901
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Fluid Discharge (2686)
Event Date 07/19/2021
Event Type  Injury  
Event Description
It was reported that after a thr was performed on (b)(6) 2021, to implant a combination of spectron stem and r3 shell, patient had discharging wound of left hip surgical site.A revision surgery took place on (b)(6) 2021 and parts were replaced with new r3 xlpe 36x54mm 20 degree liner and 36 plus 0 oxinium head.Status of patient is unknown.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no clinical/medical documentation is available.Smith and nephew has not received the explanted device and/or adequate materials to fully evaluate the complaint.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN R3 SHELL
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12277025
MDR Text Key265052060
Report Number1020279-2021-06158
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKN01100901
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/19/2021
Initial Date FDA Received08/04/2021
Supplement Dates Manufacturer Received09/10/2021
Supplement Dates FDA Received09/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-