Model Number CLV-190
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing; therefore, the root cause of the reported malfunction cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The technical support engineer was informed by the user facility that the evis exera iii xenon light source reportedly had lamp error e103 and white balance incomplete error e311 and error e931 during preparation for use.No patient involvement or harm was reported.During troubleshoot the user facility stated that the lamp life meter of the light source was at 500 hours maximum and the user facility could not white balance due to the lamp was burnt out and the spare lamp was on.The facility will change the lamp and no device will be returned.
 
Manufacturer Narrative
This supplemental report was submitted to provide additional information from the legal manufacturer.The legal manufacturer performed a review of the device history records for the concerned device and no abnormalities were found.The investigation was completed by the legal manufacturer and determined that there is no manufacturing, material or processing related cause for this failure mode.The concerned device was not returned for service/repair or to the legal manufacturer for evaluation, therefore the exact cause of the reported malfunctions could not be conclusively determined.The potential cause of the e103 error (clv lamp failed to light) and the spare lamp turned on malfunction occurred because the lamp exceeded the maximum lighting time due to the end of its service life.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12284664
MDR Text Key265400592
Report Number8010047-2021-09847
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/15/2021
Initial Date FDA Received08/05/2021
Supplement Dates Manufacturer Received09/08/2021
Supplement Dates FDA Received09/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-