• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES HEMOSPHERE INSTRUMENT; HEMOSPHERE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES HEMOSPHERE INSTRUMENT; HEMOSPHERE MONITOR Back to Search Results
Model Number HEM1
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2021
Event Type  malfunction  
Manufacturer Narrative
The product is not expected to be returned.The diagnostic log data is expected to be returned for review.When the results are available a supplemental report will be submitted.The device history record review was completed and all manufacturing inspections passed with no non-conformances.The medwatch submission numbers for the other products involved will be provided in a supplemental report when available.Current literature states there is only modest that the fick calculation is not the standard of care for estimating accurate co/ci values.There is only modest agreement between thermodilution and fick calculation estimates.Thermodilution calculation better predicts patient status and should be over fick in clinical practice.It is unknown if procedural or patient factors played a role in this event.
 
Event Description
It was reported that there were inaccurate values with the hemosphere instrument.The post-op heart surgery patient was being monitored with a hem1 and swan ganz catheter.The customer states the cardiac output values were frequently running half of the calculated fick values.It would take several hours for the hemosphere and swan values to align/correlate with calculated fick values.This occurred with two hem1 units that had been recently upgraded to be able to run flotrac.The sg catheter was switched from an upgraded hem1 to a second recently upgraded hem1 and to a third hem1 unit.All values were running with close correlation, but the co values were still up to 1l from the calculated fick value and the ci was still off by 0.5.There was no patient harm or injury.There was no inappropriate patient treatment administered.
 
Manufacturer Narrative
The product will not be returned for evaluation.The diagnostic logs were returned and reviewed.From the data it was determined that there was no defect found with the product.The submission numbers for the other products involved in this event are: 2015691-2021-04536, 2015691-2021-04537, 2015691-2021-04528, 2015691-2021-04497.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPHERE INSTRUMENT
Type of Device
HEMOSPHERE MONITOR
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key12289838
MDR Text Key265624323
Report Number2015691-2021-04532
Device Sequence Number1
Product Code DQK
UDI-Device Identifier00690103197006
UDI-Public(01)00690103197006(11)180617
Combination Product (y/n)N
PMA/PMN Number
K163381
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/18/2023
Device Model NumberHEM1
Device Catalogue NumberHEM1
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/14/2021
Initial Date FDA Received08/06/2021
Supplement Dates Manufacturer Received07/30/2021
Supplement Dates FDA Received08/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SG CATHETER
Patient Age57 YR
Patient Weight77
-
-