Model Number BR300
Device Problem Break (1069)
Patient Problem Airway Obstruction (1699)
Event Date 07/19/2021
Event Type  Injury  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer alleges that he was using item br300-the flap on the mouthpiece broke off and went into his airway.He needed the heimlich maneuver.
 
Event Description
The customer alleges that he was using item br300 the flap on the mouthpiece broke off and went into his airway.He needed the heimlich maneuver.
 
Manufacturer Narrative
After completion of investigation, it has been found that customer misuse was the main cause for this issue.Customer mouth/teeth positioning allowed for the safety valve to become detached and sequentially allowed for it to migrate into the holding chamber.The dhr was reviewed to ensure there could be no other contributing factors that would occur during production.No such contributors were uncovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MDI SPACER
Type of Device
NEBULIZER
Manufacturer (Section D)
SUNMED
2710 northridge dr. suite a
grand rapids MI 49544
MDR Report Key12295254
MDR Text Key265774991
Report Number1314417-2021-00011
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberBR300
Device Lot Number323666
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/19/2021
Initial Date FDA Received08/09/2021
Supplement Dates Manufacturer Received07/19/2021
Supplement Dates FDA Received09/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-