HEARTWARE INC. HVAD PUMP IMPLANT KIT; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1103
Device Problems Complete Blockage (1094); Device Displays Incorrect Message (2591); Noise, Audible (3273)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 06/22/2021
Event Type  Injury  
Event Description
This event report is being filed so that appropriate filing can be performed to the fda.It is a device thrombus, not related to patient noncompliance or to any mistakes made by any staff or medical team members during patient's admission to ed or in subsequent hospitalization.Patient arrived at emergency department with complaints of a grinding sound from his lvad and low-flow alarms.He had reported some recent bouts of diarrhea.Lvad coordinator downloaded the lvad logfile, which was analyzed by medtronic.Of note, there were no increases in power or increases in flow that would have indicated a thrombus.Also, inr was 2.2 and had been therapeutic for a month.No thrombus was noted per echocardiogram.A ct with contrast was deferred due to patient's renal status.Patient was admitted and put on a heparin drip.Ldh was 345 one week earlier and had risen to 455, peaking to 1522 three days later, then slowly declining to 356 two weeks later.Haptoglobin remined low at 12, and max level was 18.Patient was placed on a milrinone drip and listed as status 2 for transplant due to device malfunction.He received heart transplant two weeks after initial ed visit.Plan to return lvad to manufacturer for analysis.Manufacturer response for lvad, heartware (per site reporter).
 
Event Description
This event report is being filed so that appropriate filing can be performed to the fda.It is a device thrombus, not related to patient noncompliance or to any mistakes made by any staff or medical team members during patient's admission to ed or in subsequent hospitalization.Patient arrived at emergency department with complaints of a grinding sound from his lvad and low-flow alarms.He had reported some recent bouts of diarrhea.Lvad coordinator downloaded the lvad logfile, which was analyzed by medtronic.Of note, there were no increases in power or increases in flow that would have indicated a thrombus.Also, inr was 2.2 and had been therapeutic for a month.No thrombus was noted per echocardiogram.A ct with contrast was deferred due to patient's renal status.Patient was admitted and put on a heparin drip.Ldh was 345 one week earlier and had risen to 455, peaking to 1522 three days later, then slowly declining to 356 two weeks later.Haptoglobin remained low at 12, and max level was 18.Patient was placed on a milrinone drip and listed as status 2 for transplant due to device malfunction.He received heart transplant two weeks after initial ed visit.Plan to return lvad to manufacturer for analysis.Manufacturer response for lvad, heartware (per site reporter).Unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HVAD PUMP IMPLANT KIT
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key12300579
MDR Text Key265817247
Report Number12300579
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial,Followup
Report Date 07/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/13/2021
Event Location Hospital
Date Report to Manufacturer08/10/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/10/2021
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/12/2021
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age17885 DA