• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Model Number IPN913776
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported "temp alarm icon on heater not working".No patient involvement reported.
 
Event Description
It was reported "temp alarm icon on heater not working".No patient involvement reported.
 
Manufacturer Narrative
Qn# (b)(4).The sample was returned for evaluation.A visual exam was performed and there was a crack observed in the rear mounting bracket.There was also a piece missing from the trim.No other defects were observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was able to properly operate with no issues.The high temperature visual and audio alarms were working properly.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key12300834
MDR Text Key265811049
Report Number3003898360-2021-00676
Device Sequence Number1
Product Code BTT
UDI-Device Identifier14026704646811
UDI-Public14026704646811
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberIPN913776
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2021
Initial Date Manufacturer Received 07/20/2021
Initial Date FDA Received08/10/2021
Supplement Dates Manufacturer Received09/02/2021
Supplement Dates FDA Received09/03/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-