OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2021
Event Type  malfunction  
Manufacturer Narrative
Olympus technical assistance center (tac) support spoke to the customer to further understand the reported event.The customer was not familiar with the equipment and did not have an endoscope attached.Tac explained that an endoscope must be attached prior to use.The customer obtained a scope and the unit was working as normal and light emitted out of the distal end appropriately.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported the visera elite xenon light source has no light illumination.The event occurred during preparation for use.There was no patient involvement, no harm or user injury reported due to the event.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12307946
MDR Text Key266034686
Report Number8010047-2021-10099
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/16/2021
Initial Date FDA Received08/11/2021
Supplement Dates Manufacturer Received08/26/2021
Supplement Dates FDA Received08/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-