Catalog Number UNKAA097
Device Problems Improper or Incorrect Procedure or Method (2017); Patient-Device Incompatibility (2682)
Patient Problems Inflammation (1932); Unspecified Tissue Injury (4559)
Event Date 07/20/2021
Event Type  Injury  
Manufacturer Narrative
Based on the information provided, no conclusion can be made.As reported, postoperative usage of the arista ah hemostatic agent the patient had an increase in tissue retraction and inflammation.It was reported that the patient¿s issue resolved and no additional treatment was required.The doctor does not recall removing/irrigating any of the residual arista powder after application.The warning section of the instruction for use supplied with this device states, " once hemostasis is achieved, excess arista¿ ah should be removed from the site of application by irrigation and aspiration" and also states, "arista¿ ah swells to its maximum volume immediately upon contact with blood or other fluids.Dry, white arista¿ ah should be removed." no lot number has been provided; therefore, a review of the manufacturing records is not possible.Note, the date of event is consider to be a best estimate as (b)(6) 2021 based on date of awareness.Should additional information be provided, a supplemental mdr will be submitted.
 
Event Description
As reported, the surgeon used the bard/davol arista ah hemostatic agent during a lumpectomy case.As reported, the surgeon did not recall removing/irrigating any of the residual arista powder after application.It was reported that, after evacuating a hematoma it was noticed the patient had an increase in tissue retraction and inflammation post operatively.As reported, the patient¿s issue resolved and there was no treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARISTA AH
Type of Device
AGENT, ABSORBABLE HEMOSTATIC, NON COLLAGEN BASED
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
DAVOL INC. -1223089
160 new boston st.
woburn MA 01801
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key12316180
MDR Text Key266369331
Report Number1213643-2021-20248
Device Sequence Number1
Product Code LMG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKAA097
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/20/2021
Initial Date FDA Received08/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-