Model Number 0998-00-3023-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Testing of actual/suspected device (10/213): the getinge field service engineer (fse) that encountered the issue replaced the 12v battery and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted if this information is provided to us.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery runtime.There was no patient involvement, and no adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Analysis of production: (3331/213) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period aug 2019 through jul 2021 was reviewed.There were no triggers identified for the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key12318618
MDR Text Key266404238
Report Number2249723-2021-01801
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/21/2021
Initial Date FDA Received08/13/2021
Supplement Dates Manufacturer Received09/16/2021
Supplement Dates FDA Received09/17/2021
Patient Sequence Number1
-
-