MAKO SURGICAL CORP. HANDPIECE MICS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 209063
Device Problems Loss of Power (1475); Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2021
Event Type  malfunction  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event Description
Mako mics handpiece would not cut.Power to the handpiece but the saw would not oscillate and there was metal debris.New mics handpiece used.
 
Manufacturer Narrative
Reported event an event regarding non functional involving a mako mics was reported.The event was confirmed.Method & results product evaluation and results: handpiece mics ¿ 209063 ¿ serial# (b)(6) - rma#331104 inspected and determined failure of the following test steps: 7.1.5.1 original cable agitation test 7.1.5.3.4 tn 0835 cable functional (new cable) original description confirmed.Disposition: rtv - motor failure inspected by: spencer barfield clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices, including serial number (b)(6), were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been 1 other similar event for the lot referenced.Conclusions: the alleged failure mode was confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Event Description
Mako mics handpiece would not cut.Power to the handpiece but the saw would not oscillate and there was metal debris.New mics handpiece used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE MICS
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12325571
MDR Text Key266702103
Report Number3005985723-2021-00141
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486030193
UDI-Public00848486030193
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number209063
Device Catalogue Number209063
Device Lot Number42040219
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/22/2021
Initial Date FDA Received08/16/2021
Supplement Dates Manufacturer Received11/03/2021
Supplement Dates FDA Received11/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-