Model Number 71631116
Device Problems Break (1069); Dull, Blunt (2407)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Event Description
It was reported that a ruler has broken threads, a t-handle has broken tabs, a 12.5mm entry reamer is dull and chipped, a short hexdriver has damaged threads and a screw length sleeve is cracked at the joint.As this was noticed upon field inspection, there was not patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned 12.5mm entry reamer confirmed the device is dull and chipped at the end.This device also has excessive wear usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12.5MM ENTRY REAMER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12326700
MDR Text Key266742890
Report Number1020279-2021-06380
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010419583
UDI-Public03596010419583
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71631116
Device Catalogue Number71631116
Device Lot Number10CM18032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2021
Initial Date Manufacturer Received 08/02/2021
Initial Date FDA Received08/16/2021
Supplement Dates Manufacturer Received08/17/2021
Supplement Dates FDA Received08/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-