Model Number 221267 |
Device Problem
Contamination of Device Ingredient or Reagent (2901)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported that while using 3 bd bbl¿ chocolate ii agar (gc ii agar with hemoglobin and isovitalex¿) contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter problem: nathalie reported receiving chocolate ii agar plates containing biological contamination ".
|
|
Manufacturer Narrative
|
Investigation summary: during manufacturing of material 221267, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 1106172 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 1106172.Retention samples from batch 1106172 were not available for inspection.Three photos were received for investigation.One photo shows a log with an entry for batch 1106172 three lines.The other two photos each show the agar surface of an opened plate with bacterial growth in the media and on the surface.No plate print or product labels are visible in either photo for batch verification.Without batch verification, the photos cannot confirm the complaint.No return samples were received for investigation.A trend was identified for contamination and investigation found opportunities for bioburden reduction in the manufacturing process.A capa (corrective and preventative actions) 3076308 has been initiated and involves implementing additional cleaning events and evaluation of manufacturing procedures focused on in-process bioburden reduction.Additional trainings are planned with an ongoing training review for cleaning processes.This complaint can be confirmed by the trend for biological contamination.Bd will continue to trend complaints for contamination.
|
|
Event Description
|
It was reported that while using 3 bd bbl¿ chocolate ii agar (gc ii agar with hemoglobin and isovitalex¿) contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter problem: (b)(6) reported receiving chocolate ii agar plates containing biological contamination.
|
|